Instrument Selection

DP Transmitters: Review Differential Span and Common Static Pressure Separately

A small pressure difference can exist between two high-pressure points. Review the pressure on each side as well as their difference.

Library dates organize the collection. Actual publication and revision dates are shown separately.

A small pressure difference can exist between two high-pressure points. Review the pressure on each side as well as their difference.

Keep two specifications distinct

WIKA’s DPT-20 description lists differential measuring ranges and static-pressure limitations separately. These address different conditions. Selecting a suitable small differential range does not, by itself, establish that the instrument can withstand the pressure simultaneously present at its ports.

Use an invented two-port case

Suppose the high port is at 500 kPa and the low port at 495 kPa under a compatible reference. Their differential is 5 kPa, although both sides are near 500 kPa. A comparison sheet showing only “5 kPa required” conceals the common-pressure requirement.

Review normal and relevant transient cases for the actual arrangement, including any manufacturer limits on one-sided pressure. Do not infer those limits from a differential display scale. The invented pressures demonstrate specification separation; they do not rate a model or installation.

Request a complete application match

Give the supplier the required DP interval, pressure at each port, media and relevant conditions. Retain the exact model’s static, differential and unilateral pressure statements with the offered configuration. Performance effects under line pressure should also be checked where material to the measurement task.

Primary reference: Manufacturer technical reference.

Instrument evidence worksheet

This blank worksheet is for your own project. It contains no H M machine trial result.

Instrument evidence worksheet
Review questionEvidence to retain
What difference is required?Signed differential interval
What pressure reaches each port?Common/static and transient cases
Which limits apply?Exact model and configuration evidence

Customer Questions

Does a 5 kPa differential mean both ports are near atmosphere?

No.

Can the differential span establish the static rating?

No. Review the separate specification.

Are one-sided limits automatically the same as balanced limits?

Do not assume that. Check the actual documentation.