Pt100 or Pt1000: Match Nominal Resistance and the Configured Curve

A platinum probe label is only part of its electrical identity. Match nominal resistance, curve information and input configuration before using converted temperatures.

Library dates organize the collection. Actual publication and revision dates are shown separately.

A platinum probe label is only part of its electrical identity. Match nominal resistance, curve information and input configuration before using converted temperatures.

Keep the identifying properties together

An RTD conversion needs the applicable nominal resistance and resistance-versus-temperature relationship. Platinum material alone does not identify every coefficient. The input must support and be configured for the actual sensor relationship; a label saying “RTD” leaves that question open.

Compare two invented quotation lines

Line A specifies a 100 Ω nominal platinum element with its curve identification. Line B specifies a 1,000 Ω nominal element but copies the same controller configuration note. Ask whether that note was reviewed for line B, rather than treating both probes as interchangeable because their process connection is the same.

Next distinguish the sensor’s permitted temperature interval from the controller’s selectable display interval. A software option does not expand the physical probe’s rating. Retain the supplied certificate or specification as the source of the conversion identity.

Approve a complete configuration record

List the probe model, nominal resistance, applicable curve or coefficients, wiring arrangement and configured input type in one record. Ask for a suitable documented check of the implemented conversion. Avoid entering coefficients from an unrelated probe sheet or assuming that a new input recognizes every connected RTD automatically.

Primary reference: Manufacturer technical reference.

Instrument evidence worksheet

This blank worksheet is for your own project. It contains no H M machine trial result.

Instrument evidence worksheet
Review questionEvidence to retain
What is the sensor identity?Model and nominal resistance
Which conversion applies?Curve or coefficient reference
What is configured?Input type and supported interval

Customer Questions

Does platinum material identify the complete curve?

No. Retain the applicable conversion information.

Does a controller display range rate the probe?

No. Review probe and input limits separately.

Can the configuration note be copied between offers?

Only after confirming its applicability to the actual sensor.